specifiche
Al fine di soddisfare le esigenze specifiche delle varie applicazioni i datalogger sono disponibili in diversi modelli e tipi di sensori. Per le specifiche complete, si rimanda alle tabelle di seguito
YL-M32† MicroV Log |
YL-M33 †† V&C Log |
YL-M34 †† V&C Log |
YL-M41 Pulse Log | YL-M42 Pulse Log | YL-M61 Lux Log | YL-M62 Lux / RH / T°C |
YL-M90 †† Multi Log |
|
Sensor type | 2 x inputs | 1 x volt and current inputs | 2 x volt and current | 1 x input channel | 2 x input channels | 1 x luminance | 1 x luminance, integrated 1 x temperature and humidity | 1 x integrated humidity/temperature, 2 x additional inputs (including Pt1000 or Pt100) |
Internal temperature range / resolution | NA | -40 to +90 °C / 0.01K | ||||||
External temperature range / resolution | NA | -70 to +250 °C / 0.01K | ||||||
Temperature operating range with CR2477N battery **** |
-30 to +70°C by default * | |||||||
Temperature operating range with CR2450HT battery |
-40...+90°C | |||||||
Humidity sensor (internal & external) range/resolution | NA | 0..100%RH / 0,01%RH | ||||||
Condensation sensor | bedewed / dry | |||||||
Cable | DKC-S for ranges 0 to 1V DKC-U for ranges > 1V |
NA | ||||||
Accuracy | 0.1% of range | NA | ||||||
Meauring pulses (potential-free) / range/ resolution | NA | 0 to 65 000 pulses, interval of 0 to 100 pulses / sec. 1Hz / 1 pulse |
NA | 0 to 65,000 pulses, interval 0 to 100 pulses per sec / 1Hz,1 pulse | ||||
Voltage pulses / range/ resolution | NA | 0 to 65 000 pulses, interval 0 to 1300Hz 1Hz / 1 pulse |
NA | 0 to 65,000 pulses, interval 0 to 1300 Hz /1Hz, 1 pulse | ||||
Signal input | NA | Potential-free signals or low level pulses < 0,5 VDC and high level between 2 and 3 VDC can be processed. ** | NA | Potential-free signals or low level pulses < 0.5 VDC and high level between 2 and 3 VDC can be processed with the standard cable YY-CS (included in delivery) *** |
||||
Mesuring range Light/relative humidity/temperature | NA | 0 to 40 000 Lux / 0 to 100%RH / -30 to +70°C | NA | |||||
Dimensions d x h (mm) | 50/30 | 50/33 | 50/30 | |||||
Memory capacity/optional | 100,000 readings/ 4 million readings | |||||||
Interval | 1sec to 24h | 1 sec to 24h / Fastmode 32, 16, 8, 4, 2Hz | 1sec to 24h | 1 sec to 24h/ Fastmode 2Hz | ||||
Enclosure | Robust, shock-proof POM, protection class IP65 | |||||||
Standard lithium battery life | 4 years @ 1 min / 230 days @ 10 sec / 25 days @ 1 sec | NA | ||||||
Included in delivery | Data logger, 1 battery, YoyoView Software, USB cable, manual, 2 connection cables | Data logger, 1 battery, YoyoView Software, USB cable, connection cable YY-CS(max.5V), manual, declaration of conformity |
Data logger, 1 battery, YoyoView, USB cable, manul, declaration of conformity | |||||
Optional | YoyoView Plus Software, calibration certificate, wall holder, carrying case, lead seal kit |
YY-CP connection cable (2 m for pulses with voltages from 5 to 24V YoyoView Plus Software, calibration certificate |
YoyoView Plus, calibration certificate, lead seal kit, wall holder |
* Using the logger within the extended operating range requires the use of the CR2450HT battery.
** Higher levels up to 24 V require the YY-CP cable.
*** Higher levels require up to 24 V DKC-P cable.
**** PVC cable type V can be used from -20 to +80°C, Teflon cable type G can be used from -70 to +250°C. The humidity probe YS-RHT120 may be used only between -40 to +120°C.
†
Strain gauges measurements as bridge circuits (for full bridges 60...700Ohm) | |||||
Range mV | ± 5 | ± 10 | ± 20 | ± 50 | ± 100 |
Resolution µV * | 0.15 | 0.3 | 0.6 | 0.8 | 1.5 |
Input impendance Mohm | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 |
* Strain gauge measurements allow a sample rate of up to 32 Hz in fast mode. The maximum resolution is ten times of the above value.
Strain gauges measurements as bridge circuits (for full bridges 60...700Ohm) | ||||||
Range mV | ± 5 | ± 10 | ± 20 | ± 50 | ± 100 | ± 1000 |
Resolution µV * | 0.15 | 0.3 | 0.6 | 0.8 | 1.5 | 15 |
Input impendance Mohm | 1 | 1 | 1 | 1 | 1 | 1 |
* High impedance mode allows for a maximum sample rate of 1 Hz.
Input ranges | ||||||||
Range mV | 0 to 10 | 0 to 20 | 0 to 50 | 0 to 100 | 0 to 1V | 0 to 2.5V | 0 to 5V | 0 to 10V |
Resolution µV * | 0.58 | 0.58 | 0.76 | 1.54 | 15.4 | 38.9 | 76.9 | 154 |
Input impendance Mohm | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 0.1 | 0.1 |
* input ranges can be measured at 32 Hz. At 32 Hz the maximum resolution is ten times of the above value.
††
Input ranges | |||||
Range mV | 0 to 1V | 0 to 2.5V | 0 to 5V | 0 to 10V | 0 to 24mA |
Resolution µV * | 15.4 | 38.9 | 76.9 | 154 | 0.36µA |
Input impendance Mohm | 0.1 | 0.1 | 0.1 | 0.1 | 10 Ohmshunt |
* input ranges can be measured at 32 Hz. At 32 Hz the maximum resolution is ten times of the above value.
Note: A YY - CU cable with integrated potential divider is required in order to connect voltage signals. Current signals require the YY-CI cable.
†††
Voltage and current | ||||||||
Range mV | 0 to 10 | 0 to 20 | 0 to 50 | 0 to 100 | 0 to 1V | 0 to 2.5V | 0 to 5V | 0 to 10V |
Resolution µV * | 0.58 | 0.58 | 0.76 | 1.54 | 15.4 | 38.9 | 76.9 | 154 |
Input impendance Mohm | 2.5 | 2.5 | 2.5 | 2.5 | 2.5 | 0.1 | 0.1 | |
Voltage range/resolution | 0 to 24mA / 0.36µA | |||||||
Accuracy | 0.1% of range | |||||||
Signal input | Lower signals (max 1V) can be processed with the standard cable YY-CP. Higher voltage levels require the connection cable YY-CP. |